User Manual
22
Cleaning and Disinfection between Users
To prevent infection and eliminate possible pathogen exchange between users due to contamination, cleaning and
disinfection of the device and its accessories shall be performed by qualified personnel when used between users.
1. Remove battery and disconnect all external power from the device.
2. Dispose of and replace all accessories not suitable for multiple users including cannulas and oxygen tubing.
3. Clean all exterior surfaces using Super Sani-Cloth germicidal disposable wipes or equivalent. Remove all visible
contamination from the external surfaces of the device, battery and accessories. Be sure to closely inspect and remove
contamination from seams and recesses on the device that may trap contaminants. Wipe with clean paper towel to
remove debris.
4. After all visible contamination is removed; use a second germicidal wipe to thoroughly wet the surfaces of the device
and accessories. Allow to remain wet for 4 minutes. Use additional wipes if needed to assure surfaces are wetted
continuously for 4 minutes.
5. Allow device to air dry completely.
6. Inspect the device for visible contamination. Repeat cleaning/disinfection process if necessary.
Storage
1. Remove battery(s) prior to storage.
2. Store the device and battery(s) in a cool, dry area.
Disposal Instructions
This device may contain substances that could be harmful to the environment and must be
disposed of properly.
Follow local governing ordinances and recycling plans regarding disposal of the device and
accessories.
Troubleshooting
WARNING
Failure to resolve an alert condition may cause the device to shut down.
Technical Alerts Description
The device monitors various internal components and compares them to acceptable limits. An alert is generated when the
acceptable limit has been exceeded.
Alerts are classified as Low Priority Technical Alert Conditions. An alert requires the user to perform an action. The user is
notified of an alert condition by an audible beep every 16 seconds and flashing yellow LED light.
When an alert condition occurs, the user may press the Paused Alert button
to silence the alert and switch the LED
alert indicator from flashing to continuous for a 5 minute silence period. During this silence period, if the alert condition is
corrected, the LED alert indicator will turn off.