User Manual
20
Specifications
Dimensions: Height: 8.4 in (21.4 cm), Width: 3.2 in (8.3 cm), Depth: 8.5 in (21.6 cm)
Weight: 5.0 lbs (2.2 kg) with single battery and no carry bag
Altitude: Up to 10000 ft (3048 m) above sea level
Operating Temperature: 41˚F to 104˚F (5˚C to 40˚C)
Operating Atmospheric Pressure: 700-1060 hPa
Storage / Transport Temperature: -13˚F to 158˚F (-25˚C to 70˚C)
Operating / Storage /
Transport Conditions:
Up to 90% non-condensing relative humidity, water vapor pressures
up to 1.48 in Hg (50 hPa)
Oxygen Purity: 87% to 95.5%
Start Up Time: ≥ 87% O2 Concentration < 2 min*
Delivered Oxygen Pulse Volume:
Setting 1: 220 +/-15% mL/min
Setting 2: 440 +/-15% mL/min
Setting 3: 660 +/-15% mL/min
Setting 4: 880 +/-15% mL/min
Setting 5: 1000 +/-15% mL/min
Breathe Rate: 15-40 BPM (breaths per minute) without reduction of bolus minute volume.
Trigger Sensitivity: < -0.45 cmH2O
Power:
AC to DC Power Supply:
Input 100-240 VAC, 50-60 Hz (< 2.0 A)
Output 18 VDC (up to 5.56 A)
DC to DC Power Supply (Automotive):
Output 18 VDC (up to 6.67 A
Device Battery: 14.8 Vdc, 6.4 Ah, 94.7 Wh
Maximum Outlet Pressure: 12 psi (83 kPa)
Battery Duration (Approximate):
Setting 1: 6.5 hrs
Setting 2: 4.3 hrs
Setting 3: 2.7 hrs
Setting 4: 2.0 hrs
Setting 5: 1.5 hrs
Sound Pressure Level
(@ Setting 2):
< 40 dBA
Sound Power Level (@ Setting 2): < 48 dBA
Audible Signal Sound
Pressure Level
> 55 dBA
Applied Parts: Cannula/Oxygen Tubing, Oxygen Outlet Port, Carry Bag
Electrical Classification:
Class II Electrical Shock Protection, Type BF Applied Part, IP22 Ingress Protection
Rating, Continuous Operation
Regulatory Listings:
The Live Active Five Portable Oxygen Concentrator has been designed, tested, and
certified to the following regulatory standards.
ANSI/AAMI 60601-1 Ed 3.1
IEC 60601-1-2: 2014
CAN/CSA 22.2 No. 60601-1
ISO 80601-2-69
ISO 80601-2-67
IEC 60601-1-6
IEC 60601-1-8
IEC 60601-1-11
RTCA DO 160G
Volatile Organic Compound (VOC) and
Particulate Requirements:
The oxygen delivered from the PM4155 “Live Active Five” Portable Oxygen
Concentrator meets the following requirements for particulate levels, VOC levels,
carbon monoxide levels, carbon dioxide levels and ozone levels.
ISO 18562-2
ISO 18562-3
21 CFR 801.415
EPA NAAQS
Particulate Matter
VOC Levels
Ozone Levels
Carbon Monoxide Levels
OSHA Permissible Exposure Limits: Carbon Dioxide Levels
Standard Test Method for Determination of Volatile Organic Chemicals in Atmospheres
(Canister Sampling Methodology)
* May vary based on age of device.
Specifications are subject to change without prior notice.