User Manual
19
Warnings
This device is not to be used for support or to sustain life. This device is intended to provide supplemental oxygen only.
This device is not intended for newborn and infant use.
Precision Medical Inc, and your equipment provider are accountable for ensuring the compatibility of the device and all of the parts or
accessories used.
Use of accessories or replacement parts not listed in this Manual may cause adverse eects to basic safety or essential performance of
the device and will void warranty.
If you are unable to understand the warnings, cautions or instructions, contact a health care provider or technical personnel before
attempting to use this device.
Users with hearing and/or sight impairment(s) may need assistance while using this device.
Users who breathe from their mouths or through an oxygen mask should not use this device.
If you feel discomfort or are experiencing a medical emergency while undergoing oxygen therapy, seek medical assistance immediately
to avoid harm.
In the event of an alert condition or if you are experiencing any signs of discomfort, connect to another oxygen source. Contact your
Provider and/or Healthcare Professional immediately.
Users unable to communicate discomfort will require additional monitoring to convey the information about the discomfort and or the
medical urgency to the care giver to avoid harm.
A risk of fire is associated with the use of oxygen, and is likely to result in fire or death. Do not use the device or accessories near any
type of flame, sparks or flammable/explosive substances.
Smoking during use of oxygen therapy is dangerous and is likely to result in facial burns or death. Do not allow smoking within proximity
of the device. If you intend to smoke, turn the device o, remove the cannula and leave the room where the cannula and the device are
located.
Oxygen makes it easier for a fire to start and spread. Do not leave the nasal cannula on combustible materials such as bed coverings,
chair cushions, etc. If the device is turned on, but not in use; the oxygen will make the materials flammable. Turn the device o when not
in use to prevent oxygen enrichment.
Use of this device at an altitude above 10,000 ft (3048 m) or outside a temperature range of 41°F to 104°F (5°C to 40°C) or a relative
humidity above 90% may adversely aect the flow rate and the percentage of oxygen and consequently the quality of the therapy.
The electrical cord and/or tubing could present a tripping or strangulation hazard.
Keep away from children and pets.
The device must be used in dry conditions. Do not submerge, operate under water, bathe or swim while in use.
Wind or strong draught can adversely aect accurate delivery of oxygen therapy. Examples: Using this device beside an open window, in
front of a fan, or in the back seat of an open convertible car can aect the accuracy of oxygen delivery.
If a prescribing Healthcare Professional has determined that an interruption in the supply of oxygen, for any reason, may have serious
consequences to the user, an alternative source of oxygen should be available for immediate use.
DO NOT operate the device without the inlet filter or while that inlet filter is wet to prevent damage to the device.
The device, its parts or accessories do not contain known phthalates which are classified as carcinogenic, mutagenic or toxic.
This device should only be used when prescribed by a physician. The use of non-prescribed oxygen therapy can be hazardous.
ALWAYS confirm your prescribed flow setting before use and monitor on a frequent basis.
ALWAYS keep some distance from walls, furniture, and especially curtains that could prevent adequate airflow to the device.
ALWAYS use parts recommended by the manufacturer to ensure proper function and to avoid the risk of fire and burns.
DO NOT lubricate fittings, connections, tubing, or other accessories of the device to avoid the risk of fire and burns. Use only water-based
lotions or salves that are oxygen-compatible before and during oxygen therapy. Never use petroleum or oil-based lotions or salves.
DO NOT cover or obstruct device ventilation. The air inlets and outlets requires proper ventilation.
DO NOT disassemble or attempt to repair. There are no user serviceable parts inside. Contact Precision Medical, Inc. for service.
DO NOT modify this device.
DO NOT reach for the device if it has fallen into water. Unplug immediately if the device has fallen into water.
DO NOT use a humidifier bottle with this device.
DO NOT use while sleeping.