User Manual
15
Oxygen Sensor Replacement (508419):
*Ensure to follow Electrostatic Discharge procedures to avoid damage to electronic components
Tools and equipment required:
• #1 Phillips screwdriver
• Needle nose pliers
Disassembly
1. On back of battery separator plate (508539), remove the 2 top screws (508662) that attach the battery separator plate
(508539) to the top case (508534).
2. Carefully turn top case (508534) toward PC board (508552) and remove the 2 ribbon cables from the PC board.
Screw x2 (508662)
Battery Separator Plate (508539)
Top Case (508534)
Top Case (508534)
Ribbon Cables