Instruction Manual

6
CAUTION
The POC is designed to be used indoors or outdoors, in the outdoors
environment the POC should be used in a “dry location” with Battery only.
The proper placement and positioning of the prongs of the nasal cannula in the
nose is critical to the amount of oxygen delivered to the respiratory system of
the patient.
DO NOT place any liquids on or near the POC. If any liquid gets on the POC,
immediately turn “OFF” the power switch and unplug from the electrical outlet.
DO NOT use while sleeping.
GENERAL SPECIFICATIONS
When measuring any published tolerance, be sure to consider the tolerance of
accuracy of the measuring equipment.
Gas volume and flowrate specifications for gas delivered to the patient are
expressed at ATPD (ambient temperature and pressure, dry).
Precision Medical Inc, and your equipment provider are accountable for
ensuring the compatibility of the oxygen concentrator and all of the parts or
accessories used to connect to the patient before use.
The Oxygen Concentrator, its parts or accessories do not contain known
phthalates which are classified as carcinogenic, mutagenic or toxic.
The Oxygen Concentrator, its parts and accessories are intended for use by a
single patient and should be cleaned\disinfected before use on a new patient.
Note: The nasal cannula cannot be cleaned and should be deposed of.
SPECIFICATIONS
PM4130 PM4150
Dimensions: Height: 8.5 in (21.6 cm) 10.1 in (25.7 cm)
Width: 6.5 in (16.5 cm) 6.5 in (16.5 cm)
Depth: 4.5 in (11.4 cm) 4.5 in (11.4 cm)
Weight: 4.9 lbs (2.22 kg) 6.6 lbs (2.98 kg)
Shipping Weight: 10.8 lbs (4.91 kg) 12.15 lbs (5.6 kg)
Altitude: Sea level to 14,000 ft (4267 m)
Operating Temperature: 41˚F to 104˚F (5˚C to 40˚C)
Storage / Transport Temperature: -4˚F to 140˚F (-20˚C to 60˚C)
Operating / Storage / Transport
Relative Humidity: Up to 95% Non-Condensing
Start Up Time: 87% O2 Concentration ≤ 10 min