PM4130 PM4150 PM4150P Model: PM4100 Series PM4130P
CONTENTS RECEIVING / INSPECTION ........................................................................1 INTENDED USE............................................................................................1 READ ALL INSTRUCTIONS BEFORE USING ........................................1 EXPLANATION OF ABBREVIATIONS ...................................................2 SAFETY INFORMATION - WARNINGS AND CAUTIONS..................2 SPECIFICATIONS.................................................................
EXPLANATION OF ABBREVIATIONS mL/min Milliliters Per Minute LED Light Emitting Diode POC Portable Oxygen Concentrator Indicator light Colors and Meaning Red Warning - Immediate response by the OPERATOR is required Amber Caution - Prompt response by the OPERATOR is required Green Ready for use SAFETY INFORMATION - WARNINGS AND CAUTIONS DANGER Indicates an imminently hazardous situation which, if not avoided, will result in death or serious injury.
SAFETY INFORMATION - WARNINGS AND CAUTIONS (continued) General Warning Sign NO SMOKING Follow Instructions for Use DO NOT Disassemble General Mandatory Action Sign NO Oil or Grease The Oxygen Concentrator, gas pathways, components and accessories do not contain any natural rubber latex. Protected from touch by fingers and objects greater than 12 millimeters. \ Protected from condensation.
WARNING Use the POC for its “Intended Use” as described in this manual. This POC contains magnetic, ferrous material that may affect the results of an MRI. In the event of an equipment alarm or if you are experiencing any signs of discomfort, connect to another source of oxygen and contact your Provider and / or Health Care Professional immediately. In certain circumstances oxygen therapy can be hazardous; any user should seek medical advice prior to using this POC.
WARNING (continued) Use only spare parts recommended by Precision Medical to ensure proper function and to avoid the risk of fire and burns. Use of this device at an altitude above or below sea level to 14,000 ft (4267 m) or outside a temperature of 41°F to 104°F (5°C to 40°C) or a relative humidity above 95% Non-Condensing is expected to adversely affect the flowrate and the percentage of oxygen and consequently the quality of the therapy.
CAUTION The POC is designed to be used indoors or outdoors, in the outdoors environment the POC should be used in a “dry location” with Battery only. The proper placement and positioning of the prongs of the nasal cannula in the nose is critical to the amount of oxygen delivered to the respiratory system of the patient. DO NOT place any liquids on or near the POC. If any liquid gets on the POC, immediately turn “OFF” the power switch and unplug from the electrical outlet. DO NOT use while sleeping.
SPECIFICATIONS (continued) Oxygen Minute Volume*: PM4150 only 240 +/-15% mL/min 380 +/-15% mL/min 520 +/-15% mL/min 660 +/-15% mL/min 780 +/-15% mL/min Setting 1: Setting 2: Setting 3: Setting 4: Setting 5: Oxygen Concentration: 87% to 95% on all settings Mode of Delivery: Pulsed Dose only Trigger Sensitivity: 0.50 cmH2O maximum Trigger Delay: 0.11 seconds typical Power: Battery Duration: (Approximate) AC Power Adapter: DC Power Adapter: 100-240 VAC (1.0 A @ 120 VAC) 11-16 VDC (5.
COMPONENT DESCRIPTION CAUTION Missing or illegible labels must be replaced, contact Precision Medical, Inc. Handle Setting Knob PM4130P Setting Panel PM4150P Setting Panel Inlet Filter Power Input Jack Oxygen Outlet Fitting Internal Battery Level Indicator Setting Indicator Alarm Indicator Product Serial # Label PM4130 shown OPERATING INSTRUCTIONS WARNING Read this User Manual before operating the POC. The POC shall be used by an adult or supervised by an adult when being used by a child.
CAUTION Inspect the POC for visual damage before use, DO NOT USE if damaged. To prevent damage to the POC, DO NOT operate the POC without the inlet filter or while the inlet filter is wet. Using the EasyPulse POC 1. Verify battery status (level) prior to each use, charge as necessary. 2. Connect the end of a standard adult single lumen oxygen nasal cannula with a maximum length of 7 feet (2.1 m) to the POC’s oxygen outlet fitting.
Operating Position The POC can be operated in an upright position with or without the carry bag as shown below. PM4130 shown The POC can be operated lying on its back ONLY while in the carry bag as shown below. Using the Carry Bag The carry bag is intended to make the POC easier to transport, prevent movement due to vibration and protect the device from minor mishaps. CAUTION The carry bag is not intended to make the POC waterproof, or to protect the device from major falls or accidents/abuse.
Control panel and clear window properly aligned PM4130 shown Exhaust vents and mesh area properly aligned Proper carrying position WARNING The mesh on the carry bag must be aligned with the exhaust vents on the POC to keep the POC cool and operating properly. Be sure the exhaust vents are facing away from body while being carried. If the outside of the bag feels hot or causes any discomfort, set the POC down and check to be sure all vents are unobstructed.
POWER SUPPLY To prevent damage to the device, use only Precision Medical approved external power sources. AC Power Adapter: The AC power adapter connects the POC to a 100-240 VAC, 50/60 Hz wall outlet. The power adapter converts AC voltage to DC voltage in order to power the POC. The AC adapter will power the POC and simultaneously charge the internal battery.
DC Power Adapter: The DC power adapter connects the POC to an automobile’s 12 VDC outlet. The DC adapter will power the POC and simultaneously charge the internal battery. DC Plug Auto Accessory Plug Connection: To connect the DC Power Adapter to the POC, insert the DC plug connector into the power input jack on the POC and insert the other end into your automobile’s 12 V Accessory Jack.
Internal Battery Pack: The POC contains an internal rechargeable lithium ion battery pack. The Battery Level Indicator in the upper right corner of the control panel continuously displays the charge level of the internal battery while the POC is on. When the battery charge level drops to 15%, the last LED will turn solid red and a brief audible beep will be emitted once every minute.
ALARM / INDICATORS / TROUBLESHOOTING PM4130 AND PM4150 Green Ready for use Amber Caution - Prompt response by the OPERATOR is required Red Warning - Immediate response by the OPERATOR is required The Audible Alarm and LED Indicators are listed in the chart.
ALARM / INDICATORS / TROUBLESHOOTING PM4150 Visual Indicator/ Problem Audible Alarm Probable Cause Remedy Red alarm indicator is illuminated, unit is not running and setting numbers 1 & 3 are flashing Repetitive beep Internal battery is outside its operating temperature range 1. Turn off the POC and disconnect any power supplies 2. Check that the inlet filter is clean 3. Ensure inlet and exhaust vents are not blocked in any way 4.
MAINTENANCE / CLEANING WARNING Prior to cleaning, ensure the POC is turned off and unplug any external power sources. DO NOT disassemble or attempt to repair. There are no user serviceable parts inside. Contact your Equipment Provider for service. DO NOT spray or apply any cleaners directly onto the case. DO NOT allow any liquids to drip inside any case openings. DO NOT use harsh and/or flammable chemicals to clean the device. Perform cleaning as required. Cabinet: 1.
LIMITED WARRANTY AND LIMITATION OF LIABILITY Precision Medical, Inc. warrants that the Portable Oxygen Concentrator POC, (the Product), will be free of defects in workmanship and/or material for the following period: Portable Oxygen Concentrator Battery five (5) years from date of shipment one (1) year from date of shipment Precision Medical, Inc. is NOT responsible for normal wear and tear, or any neglect or abuse of the product.
300 Held Drive Northampton PA 18067, USA www.precisionmedical.com T: (+001) 610-262-6090 • F: (+001) 610-262-6080 ISO 13485 Certified For technical specifications call Precision Medical or visit wwwprecisionmedical.