User's Manual

Commissioning High Power Wideband Radio Head (WRH-V)
3-10 044-05251 Rev B
Where f
1
and f
2
are two TX signals entering a WRH-V. The specification for the maximum allowed PIM is -115
dBm per 100 kHz resolution bandwidth at rated output power. To test for PIM follow the steps listed in
Table 3-4.
Table 3-4 PIM Testing Procedure
Step Action
1 If BTS signals are present, continue with step 2. Otherwise, connect two signal generators to OCM
and set frequencies such that products are in relevent receive band (see equations above)
2 Set WRH-V UL gain to 60 dB and DL gain such that desired per carrier power is presented at output
port
3 Set Spectrum Analyzer to 100 kHz resolution bandwidth and view RX band at 3
rd
order product
frequency calculated in step 1
4 Loosen fastening screw in each corner on front cover of WRH-V under test, and open cover
5 Verify the noise floor specification for the Spectrum Analyzer. If the noise floor is higher than or equal
to -85 dBm/100 kHz, locate FON, disconnect UL cable from port P101 and connect
SpectrumAnalyzer to the cable using an SMA female - female connector (noise limit becomes -115
dBm + 60 dB = -55 dBm). Continue with step 7.
6 If noise floor is lower than -85 dBm/100 kHz, locate the FON card and connect the Spectrum Analyzer
to P117 (noise limit becomes -115 dBm + 60 dB - 30 dB = -85 dBm). If the FON only has three RF
ports on the left side, port P117 will not be available, so follow instructions listed in step 5.
7 Enable WRH-V TX signal. For details on using OM-Online, see OM-Online user manual
8 Using Spectrum Analyzer, measure noise level. If no noise exists above calculated limit in step 5
(-55 dBm) or step 6 (-85 dBm) above, as appropriate, then test is passed and no other action is
necessary
9 If a higher level of noise exists, disable the WRH-V TX signal and disconnect cable to DAS at bottom-
left outside corner
10 Connect low PIM load onto WRH-V Service antenna connector at bottom-left outside corner. Tighten
connectors securely
11 Remeasure noise level. If no noise exists above calculated limit in step 5 (-55 dBm) or step 6 (-85
dBm) as appropriate, then DAS has a PIM issue. Tighen all connections in DAS with torque
wrenches, replace any cripmed or severely bent cables, and retest. Repeat until all issues are solved
12 Otherwise, if noise is higher than limit, WRH-V is defective and must be replaced. See Chapter 4 for
Replacement procedures