Datasheet

Medline EXM 80
S P E C I F I C A T I O N S
M E D L I N E
w w w . p o w e r b o x . i n f o
MODEL OUTPUT CONTINUOUS CONTINUOUS PEAK OUTPUT PEAK OUTPUT
NUMBER VOLTAGE OUTPUT CURRENT OUTPUT POWER CURRENT*) POWER*)
EXM 80 5118 12V 5A 60W 5.5A 70W
EXM 80 5119 15V 4A 60W 4.7A 70W
EXM 80 5120 18V 3.6A 65W 4.2A 75W
EXM 80 5121 24V 2.9A 70W 3.3A 80W
INPUT
Nominal input range 100-240VAC.
Voltage range 90-264VAC. 85-99V with 70% output power.
Frequency 50/60Hz.
Input current 1.7A.
Inrush current Max 50A peak, cold start.
Input connector C8 cable connector or exchangeable input
connectors for Europe, US, UK and Australia.
Leakage current <100µA.
Isolation class Class II. Double isolation, type B applied part.
Zero load power consumption Approx 0,5 Watts at 110 VAC input.
OUTPUT
Power Up to 80W, see table.
Voltage See table.
Tolerance 5%.
Ripple and noise Typ. 1% p-p max (20MHz BW). 5% at <110VAC
input.
Efficiency 87-90% @ 230VAC full load.
Hold up time 35ms at 230VAC.
Start up time <2s max.
Line regulation 1% max full load. 5% at <100VAC input.
Load regulation 2% max @ 230VAC, 10-90% load change at
adapter output terminal. Output cable excluded.
Overcurrent protection Approx 125%, auto recovery.
Current limiting charact. Constant current.
Transient response 4ms max, 1ms typical.
Temperature coefficient Typ 0.04%/°C of Vout.
Output cable 2m, flexible. 1 wire cable, diameter 4mm. Internal
wires AWG16 white colour.
Output connector Barrel plug. 2.5/5.5/9.2mm, centre plus, white
colour.
Recommended connector Cliff type FC 68149, 5.5/2.5mm or FC681446 or
equivalents.
ENVIRONMENTAL
Operating temperature 0°C to +40°C.
Storage temperature –40°C to +75°C.
Humidity 5% to 95% non-condensing.
RoHS Unit complies with RoHS.
GENERAL
Switching frequency 40-80kHz.
Dimensions 118 x 45 x 70mm excl ac plugs.
Weight Max 300g including one ac plug.
Protection class Designed to meet IP50. (AC connector excluded).
Case 94V0. Polycarbonate. White colour.
MTBF >300,000 h at 25°C ambient temperature.
50 years power on.
Lifetime prediction Min 80,000 h at 22°C, 230VAC and 50% load.
Indicator “On” indicates with green LED.
Cleaning Use dry towel. Do not soak in water.
STANDARDS
Safety standards Approved according to IEC/UL60601-1 including
deviations for Europe, US & Canada by Intertek
Semko & UL. Fulfills IEC60950-1.
Safety markings S, ETL, UL & CE.
EMC standards IEC60601-1-2, IEC61204-3, EN55011 class B.
Harmonic current IEC61000-3-2.
emmissions
Voltage fluctuations IEC61000-3-3.
and flicker
ESD susceptibility IEC61000-4-2, ±6kV contact discharge,
±8kV air discharge.
Radiated susceptibility IEC61000-4-3, 3V/m.
EFT/Burst IEC61000-4-4, ±2kV on AC port, ±1kV on signal
ports.
Surge IEC61000-4-5, ±2kV common mode,
±1kV differential mode.
Conducted susceptibility IEC61000-4-6, 3V/m
Power frequency IEC61000-4-8, 3A/m.
magnetic field
Dips and interruptions IEC61000-4-11, 30% 500ms, 60% 100ms, 95%
5sec. Performance criteria A A* B.
* at 100-160 VAC nominal input voltage performance criteria B
*) Max duration 30 sec. per every 10 minutes.
Four international AC plugs included.
AC cable can be delivered as option
Main standard in all
European countries.
AC PLUGS MEDLINE EXM 80
Main standard in UK
Main standard in US
Main standard in
Australia, New Zealand
Specifications are subject to change without notice.