User manual
Poseidon MKVI User Manual Appendix 1
Page 78
T# Time (sec) Description Error Code Solution
53 Oxygen Sensor Calibration. Like the Positive Pressure Loop
Test (t49), this test check several things besides performing an
oxygen sensor calibration, including the composition of the
diluent and oxygen supplies, the proper funcioning of both the
oxygen and diluent calibration solenoid valves, and other
parameters associated with the oxygen sensor behavior. The
mouthpiece must remain in the CC position for the duration of
this test.
120 66=Diluent
FO2 Bad
71=Oxygen
FO2 Bad
1) Ensure Diluent and Oxygen cylinders have the correct oxygen percent-
age; 2) Make sure that you have connected the oxygen- and diluent tanks to
the correct connection on the pneumatic block; 3) Replace Primary oxygen
sensor with a known good sensor; 4) Ensure CO
2
absorbent cartridge is
installed correctly; 5) Ensure breathing loop temperature is within range
limits; 6) If test failure persist, contact an authorized Poseidon Service Center
for repair.
1) Ensure Diluent has correct oxygen percentage; 2) If test failure persists,
contact an authorized Poseidon Service Center for repair.
51 Helium Fraction Conrmation test.
This test will be run if a rig is enabled for decompression or if
the diluent conguration is not air. It will display the assumed
fraction of Helium in the diluent.
120 0=time out 1)Ensure to conrm fraction via wet switch wet procedure. 2) If test failure
persists, contact an authorized Poseidon Service Center for repair.
67=Primary
Dil. Low
68=Primary
Dil. High
72=Primary
O2 Low
73=Primary
O2 High
76=Bad Time
Constant
1) Ensure Diluent and Oxygen cylinders have the correct oxygen percent-
age; 2) Make sure that you have connected the oxygen- and diluent tanks to
the correct connection on the pneumatic block; 3) Replace Primary oxygen
sensor with a known good sensor; 4) Ensure CO
2
absorbent cartridge is
installed correctly; 5) Ensure breathing loop temperature is within range
limits; 6) If test failure persist, contact an authorized Poseidon Service Center
for repair.
52 Oxygen Fraction Conrmation test. Description; This test will
be run if a rig is enabled for decompression or if the diluent
conguration is not air. It will display the assumed fraction of
Oxygen in the diluent.
120 0=time out 1)Ensure to conrm fraction via wet switch wet procedure. 2) If test failure
persists, contact an authorized Poseidon Service Center for repair.