Datasheet

USA: Sales: 800-490-2361
Technical Support: technicalsupport@pomonatest.com
Fax: 425-446-5844
Europe: 31-(0) 40 2675 150 International: 425-446-5500
Where to Buy: www.pomonaelectronics.com
All dimensions are in inches. Tolerances (except noted): .xx = ±.02” (,51 mm),
.xxx = ± .005” (,127 mm). All specifications are to the latest revisions.
Specifications are subject to change without notice. Registered trademarks are
the property of their respective companies.
Copyright © 2014 D1094536 Rev. 101 Page 2 of 2
Model 6086
ESD Wrist Strap Tester
Calibration
Calibration of Pomona model 6086 is recommended annually. Calibration should be performed to ensure that the
tester is operating within limits. The calibration may be checked with discrete resistors.
Resistance Value (+/- 10%)
Test Display
675 KOhms
Red (Low-hazard)
825 KOhms
Green (OK-in limits)
8.5 MOhms
Green (OK-in limits)
11.5 MOhms
Red (High resistance)
To calibrate, simply place each resistance value (see table) between the touch plate and banana grounds. Observe
that the proper test display LED is on (see table) for the selected resistor. Be sure to hold the test cord at an insulated
point, so the resistance value is not affected by the body. There are no adjustments inside.