8.2 Troubleshooting guide
Issue
Possible root cause
Corrective action
No ignition or the
source turns off
Too low gas flow, or
no gas flow
Change the empty gas cylinder,
Increase the input pressure before the
regulating valve / flowmeter,
Make sure the gas valve cable is plugged
Too high pressure
Check for air leak in the vacuum chamber
Perform a turbopump maintenance
Set a lower gas flowrate setpoint
No HF power
Make sure the interlock is closed
Dirty surfaces,
unstable ion source
operating point, too
much acceleration
voltage
Perform a high voltage conditionning
sequence
reduce the voltages
No high voltage
Interlock open or
shortcut to ground
Make sure the interlock is closed
check for high voltage overcurrent
Current drift
Unstable ion source
operating point
Change the operating pressure
Change the operating UHF power
Current jumps
Not enough UFH
power, not enough
gas
Increase the UHF power
Increase the gas flow rate
Beam current is too
low
Not enough UFH
power, not enough
gas, voltage
difference between
beam and focus is
too low
Increase the UHF power
Increase the gas flowrate
Lower the focus voltage compared to the
beam voltage (maximum allowed
difference = 5kV. Will make the ion beam
more divergent)
Pressure reading does
not change when
changing the valve
setpoint
Valve not connected;
rack not powered.
Make sure that the D-Sub 9 cable the
rack and the gas regulating valve is
properly connected
Make sure that the rack is powered,
and that its power switch is on
Message ‘serial
connection failed
when lauching the IE
gun Labview
application program
RS232 cable not
connected, I/O
board freeze,
Labview freeze.
Make sure that the RS232 cable
between the rack and the PC is
properly connected
Close the program, disconnect and
reconnect the serial and restart the
program