User's Manual

TES-63 Source
User manual version 06/2021
Page 37/46
Polygon Physics SAS au capital social de 72 000 euros - 19, rue de Sassenage - 38600 Fontaine
SIREN 799 584 453 RCS Grenoble - TVA intracommunautaire FR 19 799584453 - APE 2899B
setpoints.
Description of the list returned by <get source><cr>
This command returns a list whose elements are:
1.
rack_state
2.
hf_state
3.
beam_state
4.
focus_state
5.
beampolar_state
6.
focuspolar_state
7.
de_state
8.
state7 (Not connected)
9.
state8 (Not connected)
10.
HF
11.
beam
12.
focus
13.
imaxbeam
14.
imaxfocus
15.
gasflow
16.
ao1 (data enter 2, pin 1)
17.
ao2 (data enter 2, pin 2)
7.7 Measuring a value with <meas> <devicename><cr>
Command
Description
Comments
meas hf_state
Returns the measurement
of the HF generator state.
meas beam_state
Returns the measurement
of the beam power supply
state.
meas focus_state
Returns the measurement
of the focus power supply
state.
meas beampolar_state
Returns the measurement
of the beam power supply
polarity state.
0 = positive polarity
1 = negative polarity
meas focuspolar_state
Returns the measurement
of the focus power supply
polarity state.
0 = positive polarity
1 = negative polarity