User's Manual

TES-63 Source
User manual version 06/2021
Page 34/46
Polygon Physics SAS au capital social de 72 000 euros - 19, rue de Sassenage - 38600 Fontaine
SIREN 799 584 453 RCS Grenoble - TVA intracommunautaire FR 19 799584453 - APE 2899B
7.4 Changing a setpoint with <set> <devicename> <value><cr>
Command
Description
Comments
set hf <value>
Set the HF power in Watt.
<value> must be in the range
[0,20].
Resolution = 1 milliWatt.
Ex: set hf 3.654 set HF power to
3.654 Watt.
set beam <value>
Set the beam high voltage
in kiloVolt.
<value> must be in the range
[0,15].
Resolution = 1 volt.
Ex: set beam 8.231 set the beam
high voltage to 8.231 kiloVolt
set imaxbeam <value>
Set the maximum current
delivered by the HV beam
power supply in
microAmpere.
<value> must be in the range
[0,800].
Resolution = 0.1 microAmpere.
set focus <value>
Set the focus high voltage
in kiloVolt.
<value> must be in the range
[0,15].
Resolution = 1 volt.
Ex: set focus 8.231 set the focus
voltage to 8.231 kV
set imaxfocus <value>
Set the maximum current
delivered by the HV focus
power supply in
microAmpere.
<value> must be in the range
[0,800].
Resolution = 0.1 microAmpere.
set gasflow <value>
Set the gas valve opening
in volt.
<value> must be in the range
[0,10].
Resolution = 1 milliVolt
set hfpulse 1
Set the HF power to its
maximum (about 20W) for
1 second.
set network
<value1>/<value2>
Allows modifying the IP
address and subnet mask.
<value1> is an IP4 address
<value2> is the subnet mask
Ex: set network
192.168.0.101/255.255.255.0
set totaltime 0
Set the total time of use of
the ion source to 0.