User's Manual
TES-63 Source
User manual – version 06/2021
Page 17/46
Polygon Physics SAS au capital social de 72 000 euros - 19, rue de Sassenage - 38600 Fontaine
SIREN 799 584 453 RCS Grenoble - TVA intracommunautaire FR 19 799584453 - APE 2899B
ensured by the provided SMA extension cable.
First, install the 90° SMA connector on the UHF power CF16 source flange.
Then, connect one side of the cable to the SMA 90° connector on the source flange and the other side
on the SMA connector labelled "Microwave Power" on the back panel of the control rack.
Recommended torque: 1 Nm.
CAUTION
Never operate the rack without the SMA extension cable connected between the rack
and the source flange (it could damage the UHF power generator).
e) High Voltage connections
Connect the cables to the corresponding feedthroughs. Make sure the bayonet mount is properly
locked.
f) RS232 connection
Install a crossover serial cable between the connector "serial" on the control rack and the computer.
g) RJ45 connection
Install a standard RJ45 cable between the connector "RJ45" on the control rack and the computer.
Install the 90° SMA
connector on the
flange. Then connect
the SMA cable on it.