TRAX User Manual  Page 43 
kBaudRate 
14 
UInt8 
4 – 2400 
5 – 3600 
6 – 4800 
7 – 7200 
8 – 9600 
9 – 14400 
10 – 19200 
11 – 28800 
12 – 38400 
13 – 57600 
14 - 115200 
12 
kMilOut 
15 
Boolean 
True or False 
False 
kHPRDuringCal 
16 
Boolean 
True or False 
True 
kMagCoeffSet 
18 
UInt32 
0 - 7 
0 
kAccelCoeffSet 
19 
UInt32 
0 - 7 
0 
Configuration parameters and settings for kSetConfig: 
kDeclination (Config. ID 1d) 
This  sets  the  declination  angle  to  determine  True  North  heading.    Positive 
declination is easterly declination and negative is westerly declination. This is not 
applied unless kTrueNorth is set to TRUE. 
kTrueNorth (Config. ID 2
d
) 
Flag  to  set  compass  heading  output  to  true  north  heading  by  adding  the 
declination angle to the magnetic north heading. 
kBigEndian (Config. ID 6
d
) 
Sets the Endianness of packets. TRUE is Big Endian. FALSE is Little Endian. 
kMountingRef (Config. ID 10
d
) 
This sets the reference orientation for the TRAX.  Please refer to Figure 4-2 for 
additional information. 
kUserCalNumPoints (Config. ID 12
d
) 
The user must select the number of points to take during a calibration.  Table 7-4 
provides the “Minimum Recommended” number of sample points, as well as the 
full  “Allowable  Range”.   The  “Minimum  Recommended”  number of  samples 
normally is sufficient to meet the TRAX’s heading accuracy specification, while 
less  than  this  may  make  it  difficult  to  meet  specification.    See  Section 5  for 
additional information. 