Owner manual
PNI Sensor Corporation    DOC#1016505 r06 
TRAX User Manual  Page 19 
vector, which in turn will result in an inaccurate heading reading. For this reason, the TRAX 
should be stationary when taking a measurement. 
As previously mentioned, PNI calibrates the accelerometer in its factory prior to shipment. 
But  over time  the bias and offset  of the  accelerometer  will drift.   For  this reason  PNI 
recommends the accelerometer be recalibrated every 6 to 12 months.  The user may return 
the  TRAX  to  PNI  for  accelerometer  calibration,  or  the  user  may  perform  a  user 
accelerometer calibration. The remainder of this section covers the user accelerometer 
calibration. 
5.2.1  Accel-Only Calibration 
The requirements for a good user accelerometer calibration differ significantly from the 
requirements  for  a  good  magnetic  calibration.    Specifically,  a  good  accelerometer 
calibration involves the  TRAX experiencing a  wide range of  pitch and  roll  values, 
preferably seeing both ±180° of pitch and ±180°  of roll.   Also, it is  necessary for the 
TRAX to be very still during an accelerometer calibration.  If possible, PNI recommends 
using a fixture to hold the device during calibration, although resting the TRAX on a hard 
surface normally is sufficient. 
The accelerometer either can be calibrated while mounted in the host system or it may be 
removed and calibrated outside the system.  The advantage of performing the calibration 
while mounted in the host system is the user does not need to remove the TRAX from the 
system, which can be burdensome, and a simultaneous Mag-and-Accel Calibration may 
be appropriate.  The advantage of performing the calibration outside of the system is it 
may be much simpler to obtain the desired range of pitch and roll. 
Figure  5-2  shows  the  two  basic  starting  positions  for  the  recommended  18-point 
calibration pattern.  Starting with the TRAX as shown on the left in Figure 5-2, rotate the 
device  about  its  z  axis  such  that  it  sits  on  each  of  its  4  edges,  taking  one  calibration 
sample on each edge.  Then place the TRAX flat on the surface and take a calibration 
sample, then flip it over (roll it 180°) and take another sample.  Next, starting with the 
TRAX as shown on the right, take a calibration point with it being vertical (0°).  Now tilt 
the TRAX back 45° and take another calibration point (+45°), then tilt the device forward 
45° and take another calibration point (-45°).  Repeat this 3-point calibration process for 
the TRAX with it resting on each of its 4 corners.  Note that it is possible to perform an 
Accel-Only Calibration with as few as 12 sample points, although it generally is more 
difficult to  obtain  a  good  calibration  with  just  12  sample points.    Also,  the  maximum 
number of calibration points is 18. 










