12 Error and Warning Messages
Error and warning messages Cause Troubleshooting
Electronic error! Restart the
device. If the error occurs again
switch off device and call ser-
vice!
Malfunctions in the electronic
system.
Restart the device. If the error occurs again, contact
technical service.
Calibration error! The device
must be re-calibrated. Call ser-
vice!
The device is not calibrated
properly.
The device must be re-calibrated. Contact technical
service.
Device temperature error! Turn
device off!
Temperature in device is too low
or too high.
Serious fault. Do not use device. Contact technical ser-
vice.
Make sure the device has been acclimatized to its en-
vironment before turning on.
Device contaminated! Fluid has
penetrated the device. Contact
technical service!
The device is contaminated with
fluid.
The device must be checked by an authorized service
technician. It is to be clearly marked and wrapped
twice in safety foil before sending to the service de-
partment for repairs.
Device service is almost due!
Contact Stryker for service.
Maintenance is due in 4weeks.
Continued use of the device is possible. However, the
device must be serviced by an authorized service tech-
nician soon.
Device service is due! Please
send to Stryker for service. Con-
tinue?
Last maintenance is more than
24 months ago. Flow rate and
pressure might not be accurate
anymore.
The device must be serviced by an authorized service
technician.
Excessive overpressure!
The actual pressure >45mmHg.
Determine the cause for exceeding the nominal pres-
sure. If the venting valve is disabled, the overpressure
must be reduced manually by opening the trocar.
Maximum pressure reached!
Nominal pressure limit
20/30mmHg has been reached/
exceeded.
Reduce the abdominal pressure by reducing the set
pressure or opening a stop cock.
Overpressure!
Abdominal pressure exceeds set
pressure and cannot be reduced
by the automatic venting system
within 5seconds.
Determine the cause for exceeding the set pressure. If
the venting valve is disabled, the overpressure must
be reduced manually by opening the trocar.
Gas heating - Over temperature!
Insufflation paused.
The temperature of the gas ex-
ceeds 43 °C/109 °F. The device
pauses gas heating and insuffla-
tion.
There are three options to continue insufflation:
• Wait until the temperature has decreased below
43°C, insufflation and heating will resume automatic-
ally shortly.
• Disconnect and reconnect tube set if temperature is
below 43°C insufflation with heating can be resumed
(manual start).
• Exchange tube set with a new one. If this error mes-
sage is displayed again, have the gas heating checked
by a qualified service technician.
Safety Feature Warning
Insufflation paused due to gas
exceeding 43°C/109°F. Insuffla-
tion will resume shortly.
Exceed Safety Threshold?
Actual pressure >12/15mmHg.
Actual flow >5l/min (Pediatric).
Actual flow > 6 l/min (Vessel
Harvest).
Confirm the query to further increase the pressure. Ex-
ceeding this safety limit is to be decided by and the re-
sponsibility of the user/operator.
Change gas tank!
Low insufflation gas supply. Check gas supply.
Gas level low. Prepare to change
gas tank.
Gas supply <20bar/290.1psi. Prepare for changing the gas bottle.
Check gas supply! Insufflation
paused!
Insufficient insufflation gas sup-
ply (bottle or central gas supply)
during insufflation.
Insufflation starts automatically after restoring the
gas supply.
Tube set invalid
The tube set is not valid. Insert a valid tube set.
en