Users Manual

Electromagnetic compatibility
64 / 77
IMMUNITY to proximity fields from RF wireless communications equipment
Test method IEC 61000-4-3
Test frequency
(MHz)
Band
(MHz)
Service Modulation Maximum power
(W)
Distance
(m)
IMMUNITY TEST LEVEL (V/m)
385 380-390 TETRA 400
Pulse modu-
lation
18Hz
1.8 0.3 27
450 430-470
GMRS 460, FRS
460
FM
±5kHz devi-
ation
1kHz sinus
2 0.3 28
710
704-787 LTE Band 13, 17
Pulse modu-
lation
217 Hz
0.2 0.3 9745
780
810 800-960
GSM 800/900,
TETRA 800, IDEN
820, CDMA 850,
LTE Band 5
Pulse modu-
lation
18 Hz
2 0.3 28
870
930
1720
1700-1990
GSM 1800;
CDMA 1900;
GSM 1900; DECT;
LTE Band 1, 3, 4,
25; UMTS
Pulse modu-
lation
217 Hz
2 0.3 28
1845
1970
2450 2400-2570
Bluetooth, WLAN
802.11 b/g/n,
RFID 2450, LTE
Band 7
Pulse modu-
lation
217 Hz
2 0.3 28
5240
5100-5800
WLAN
WLAN 802.11 a/n
Pulse modu-
lation
217 Hz
0.2 0.3 95500
5785
WARNING!
Portable HF communication equipment
Portable HF communication equipment can affect the performance characteristics
of the device PNEUMOCLEAR™
. Such equipment must therefore comply with a min-
imum distance of 30 cm (regardless of all calculations) from the device
PNEUMOCLEAR™
, its components and cables.
en
DRAFT