Users Manual
Annual Inspection
61 / 77
3. Connect a manometer (1) and an open Veress cannula (3) to the insufflation
tube connection (2)
.
4. Start insufflation:
Press the
START
button. The manometer registers a pulsing pressure increase.
When the pressure stabilizes, the manometer registers a pressure between 55
and 65mmHg
5. Stop insufflation:
Press the
STOP
button.
10.7 Gas Flow Rate Test
Test setup with open connection, without connected insufflation tube.
• Select a nominal gas flow rate of 15l/min.
• Start insufflation:
Press the
START
button.
• Press the gas volume button (0.0l must be displayed).
• Now start measuring for one minute.
• Stop the insufflation after one minute: Press the
STOP
button.
The gas consumption should be at least 11-12l.
Each successfully conducted test must be documented with the test log.