Users Manual
Annual Inspection
59 / 77
7. Stop insufflation: Press the
STOP
button.
The following values are displayed:
Actual pressure 0.0(mmHg)
Gas consumption >0.0(l)
8. Press the gas volume button.
Gas consumption 0.0(l)
The basic function check of the device is complete.
10.3 Pressure Sensor Test (Standard Operating Mode)
Fig.10–1Pressure sensor test setup
(1) Manometer
(2) Syringe
(3) PNEUMOCLEAR™ HIGH-FLOW
TUBE SET
(4) Silicone tube
(5) T adapter
(1)
(2)
(3)
(5)
(4)
(3)
CAUTION!
Gas extraction
Never use the syringe to extract gas from the device.
1. Set the Standard operating mode.
2. Select a nominal gas flow rate of 1.0l/min.
Do
not
press the
START/STOP
button.
3. Connect a manometer and an air-filled syringe to the insufflation tube connec-
tion.
4. Use the syringe to generate a pressure of 10mmHg, which is indicated on the
manometer.
Actual pressure display: 10(mmHg)
5. Use the syringe to generate a pressure of 20mmHg, which is indicated on the
manometer.
Actual pressure display: 20(mmHg)
6. Use the syringe to generate a pressure of 30mmHg, which is indicated on the
manometer.
Actual pressure display: 30(mmHg)