Users Manual

Annual Inspection
59 / 77
7. Stop insufflation: Press the
STOP
button.
The following values are displayed:
Actual pressure 0.0(mmHg)
Gas consumption >0.0(l)
8. Press the gas volume button.
Gas consumption 0.0(l)
The basic function check of the device is complete.
10.3 Pressure Sensor Test (Standard Operating Mode)
Fig.10–1Pressure sensor test setup
(1) Manometer
(2) Syringe
(3) PNEUMOCLEAR™ HIGH-FLOW
TUBE SET
(4) Silicone tube
(5) T adapter
(1)
(2)
(3)
(5)
(4)
(3)
CAUTION!
Gas extraction
Never use the syringe to extract gas from the device.
1. Set the Standard operating mode.
2. Select a nominal gas flow rate of 1.0l/min.
Do
not
press the
START/STOP
button.
3. Connect a manometer and an air-filled syringe to the insufflation tube connec-
tion.
4. Use the syringe to generate a pressure of 10mmHg, which is indicated on the
manometer.
Actual pressure display: 10(mmHg)
5. Use the syringe to generate a pressure of 20mmHg, which is indicated on the
manometer.
Actual pressure display: 20(mmHg)
6. Use the syringe to generate a pressure of 30mmHg, which is indicated on the
manometer.
Actual pressure display: 30(mmHg)
en