Users Manual
Annual Inspection
58 / 77
10 Annual Inspection
Measured values and tolerances
Each test conducted must be documented with date and signature on the test log.
The following measuring tools and resources are required to determine the listed
measurements and tolerances:
Manometer Range 0-100mmHg, error class 1 (±1% of span)
Syringe 60ml
PNEUMOCLEAR™ HIGH-FLOW TUBE SET
Silicone tube 4mm x 40cm
T-adapter 6/4/6
Veress cannula length 100mm
opening diameter 1.4mm,
inner cannula diameter 1.6mm
An authorized service technician must check the device if the specified parameters
and tolerances are exceeded.
10.1 Electrical Safety Test
1. Perform a visual inspection. Make sure that
– the fuse corresponds with the specifications indicated by the manufac-
turer,
– labels and stickers on device are legible,
– the mechanical condition of the device allows for its safe use,
– the device is clean to ensure proper and safe functionality.
2. Carry out the measurements for the ground leakage current, short-circuit cur-
rent/housing leakage current, and the protective conductor resistance as per
IEC62353 in the current version or according to the applicable national stand-
ard.
10.2 Basic Function Test (in Standard Operating Mode)
1. Remove insufflation tube from the device.
2. Use the On/Off switch to turn the device on. The device now conducts a device
check. An audible signal can be heard. Set the Standard operating mode.
3. A nominal pressure of 15mmHg and a nominal flow of 3l/min are set at the
factory.
4. The following values are displayed:
Nominal pressure 15*(mmHg)
Nominal gas flow 3*(l/min)
Actual pressure 0(mmHg)
Gas consumption 0.0(l)
*These values correspond with the factory setting. If values in the user menu
were changed, these changed values are displayed. Insert an PNEUMOCLEAR™
HIGH-FLOW TUBE SET.
5. Start insufflation: Press the
START
button. The following values are displayed:
Actual pressure 0(mmHg)
Veress Insufflation
is displayed. Streaming gas can be heard at the insufflation
tube.
6. Select a nominal gas flow of 20l/min. The following value is displayed:
Nominal gas flow 20l/min
Actual pressure 0(mmHg)
Insufflation Active
is displayed. Streaming gas can be heard at the insufflation
tube.
en
DRAFT