Operating the Device - General
40 / 77
5.15 Desufflation
The Desufflation function allows for the remaining CO
2
in the cavity to be removed
to a pressure of approximately 3mmHg after insufflation has been stopped. Desuf-
flation is available after the insufflation when a smoke evacuation tube is connec-
ted. After stopping, a query is displayed asking whether the desufflation function
should be started.
Fig.5–7Confirming or denying desuf-
flation
The display shows
Desufflation active
. Desufflation can be stopped at any time
with the
Cancel
button.
Desufflation ends automatically once the actual pressure in the cavity is
≤3mmHg. The display shows
Desufflation finished
.
Mode Desufflation
High Flow/Bariatric Mode available
Standard Mode available
Pediatric Mode not available
Vessel Harvest Mode not available
TAMIS Mode not available
Advanced Flow Mode available