5.11 Setting the Nominal Flow - All Operating Modes
To set the nominal flow, press the + or - buttons (see Fig. Screen displays [}28]
(10)/(12)) on the display below the actual gas flow display.
With every touch of the + or - button the nominal flow is increased/reduced in
steps of 1l/min. When working within the range of 0.1 to 2l/min in Pediatric
Operating Mode, the value is increased/reduced by 0.1 l/min. When pressing
the flow button longer, scrolling through the flow levels is activated but only
up to the respective, operating mode-dependent safety threshold (see below).
When setting the nominal flow rate, Pediatric operating mode has a safety
threshold at 5l/min, Vessel Harvest at 6l/min. If the flow is to be increased
above these values, the depicted query must be confirmed accordingly.
WARNING!
Exceeding the safety limit
Exceeding this safety limit is to be decided by and the responsibility of the user/op-
erator.
Operating Mode Safety threshold
High Flow/Bariatric Operating Mode n.a
Standard Operating Mode n.a
Pediatric Operating Mode 5 l/min
Vessel Harvest Operating Mode 6 l/min
TAMIS Operating Mode n.a
Advanced Flow Operating Mode n.a
CAUTION!
Safety threshold
The safety threshold of the nominal flow is deactivated per default.
The safety threshold of the nominal flow can be activated/deactivated in the User
Menu.
Veress insufflationVeress insufflation
is a gentle type of insufflation that prevents exceeding the pre-
set nominal pressure by the actual pressure even in case of small volumes. To min-
imize the risks in case of a faulty incision, the manufacturer recommends using
Veress insufflation
to start a procedure (filling abdomen with CO
2
).
CAUTION!
Venting system
Please note that the automatic venting system is not active during Veress insuffla-
tion.
How to activate Veress Insufflation
The activation of the Veress insufflation depends on the chosen mode.
Pediatric Mode
: With a nominal flow setting of ≤1l/min,
Veress Insufflation
is dis-
played after pressing the START/STOP button.
High Flow/Bariatric
,
Standard
,
Advanced Flow Mode
: With a nominal flow setting
of ≤5l/min,
Veress Insufflation
is displayed after pressing the START/STOP button.
Vessel Harvest, TAMIS
: Veress insufflation is not available.
5.12 Running Function Tests before Device Use
Preparation
Check all the sterile disposable items before removing them from the package to
ensure that the packaging is intact and that the expiration date is still valid.
For your own safety and that of your patient, use only original accessories (see
Chapter Accessories [
}69]).
en