Users Manual
Operating the Device - General
36 / 77
5.8 Unlocking Operating Modes
The device can be equipped with up to 6 procedure modes.
Device Standard Version Device Plus Version
High Flow/Bariatric Operating Mode High Flow/Bariatric Operating Mode
Standard Operating Mode Standard Operating Mode
Pediatric Operating Mode Pediatric Operating Mode
Vessel Harvest Operating Mode Vessel Harvest Operating Mode
TAMIS Operating Mode
Advanced Flow Operating Mode
TAMIS and Advanced Flow operating modes can be purchased separately. Contact
your sales representative for more information on how to purchase them and how
to get your device upgraded.
5.9 Displaying/Selecting Insufflation Operating Mode
The display depicts an overview of the insufflation operating modes. Unavailable
operating modes are depicted in gray and cannot be selected. Press the respective
button to select the desired operating mode (e.g. Standard).
The displayed parameters correspond with the factory settings or the values set in
the user menu (see Chapter Menu (Overview) [
}51]).
5.10 Setting the Nominal Pressure - All Operating Modes
To set the nominal pressure, press the
+
or
-
button Screen displays [}28] on the
display below the actual pressure display.
• The nominal pressure can be increased or decreased during insufflation or
while insufflation is stopped.
• With every touch of the + or - button the nominal pressure is increased/re-
duced in steps of 1 mmHg. When pressing the pressure button longer,
scrolling is activated but only up to the respective, operating mode-dependent
safety threshold (see below).
• An operating mode-dependent safety threshold (12 or 15mmHg) applies when
setting the nominal pressure. If the pressure is to be increased above these val-
ues, the depicted query on the operating screen of the insufflator must be con-
firmed accordingly.
WARNING!
Exceeding the safety limit
Exceeding this safety limit is to be decided by and the responsibility of the user/op-
erator.
Operating Mode Safety threshold
High Flow/Bariatric Operating Mode 15 mmHg
Standard Operating Mode 15 mmHg
Pediatric Operating Mode 12 mmHg/15 mmHg
Vessel Harvest Operating Mode 12 mmHg/15 mmHg
TAMIS Operating Mode 15 mmHg
Advanced Flow Operating Mode 15 mmHg
en
DRAFT