Users Manual

Operating the Device - General
34 / 77
Fig.5–5Insufflation gas humidification
indicator
(5) Humidification icon (blue)
(5)
Refilling
NOTE!
To refill the tube set
The humidification tube set provides approximately 200l of humidified gas when
filled with water or saline. After 200l of gas have been consumed, the tube set can
still be used, however, in order to maintain a high level of humidity, the tube set
should be refilled every 200 to 300l of consumed gas. To refill the tube set, press
the humidification icon and follow the instructions.
Press the humidification icon to start the refilling process. Pressing the humidifica-
tion icon will give the user the option to pause the running insufflation for refilling
purpose. While insufflation is not running tube set refilled question appears on the
display.
CAUTION!
Filling the tube set
Filling the tube set with more than 10 ml of water can lead to formation of water
drops. Do not refill the tube set if the absorber is still soaked with water.
Humidification icons Description Information
(blue)
Humidification tube set connected, push icon
to start refill work flow.
Indicates that a tube set with humidification
ability is connected.
(grey)
Humidification tube set is not connected or
function defective and therefore deactivated.
Humidification not available.
en
DRAFT