Users Manual
Operating the Device - General
32 / 77
To enable this function, insert a valid tube set and then press the
Smoke evacu-
ation
button (11). Use the
+
/
-
buttons (see Fig. Smoke evacuation indicator [}31]
(10)/(12)) to increase or reduce the extraction intensity. The smoke evacuation
rate can be manually set to the following rates: 1, 2, 3, and 4.
Not all rates are available in all operating modes.
Nominal flow setting must be higher than smoke evacuation flow rate otherwise
actual evacuation performance will be limited.
Fig.5–3Connections of the tube set
with smoke evacuation
at insufflation trocar
at second trocar
water trap
Suggested minimum flow settings to ensure proper smoke evacuation:
Smoke evacuation rate Approximate evacuation
flow (l/min)
Minimum nominal flow (l/
min)
1 3 8
2 6 11
3 9 14
4 12 17
Smoke evacuation is not available for Vessel Harvest mode.
CAUTION!
Do not lift the water trap.
If the water trap hangs horizontally or upside down, smoke evacuation filter can be
clogged.
Smoke evacuation icons Description Information
(white)
Smoke evacuation function is
turned off.
This icon means that the smoke evacuation function is
turned off. To activate this function, press this icon.
(green)
Smoke evacuation function is
turned on.
This icon means that the smoke evacuation function is
turned on. To switch off this function, press this icon.
en
DRAFT