Users Manual

Operating the Device - General
31 / 77
Icon Description Information
(green)
Gas heater OK Gas heating function is available.
(red)
Safety Feature Warning
Insufflation paused due to gas
exceeding 43 °C/109 °F. Insuf-
flation will resume shortly.
Insufflation is automatically stopped and the heating function is deac-
tivated. There are three options to continue insufflation:
Wait until the temperature has decreased below 43°C, insufflation
and heating will resume automatically.
Disconnect and reconnect tube set. If temperature is below 43°C
insufflation with heating can be resumed (manual start).
Exchange tube set with a new one.
(red)
Gas heater defective
Insufflation can be continued in this case but the heating function is
not available.
Restart the device if you want to use the heating function. If the gas
heating is still defective, have an authorized service technician check or
repair the device.
5.5 Using the Smoke Evacuation Function
WARNING!
Surgical smoke in operating room
Do not vent surgical smoke to the operating room. Use the provided smoke evacu-
ation function with integrated filter if possible.
WARNING!
Increased concentration of CO
2
in the operating room
Smoke evacuation function may increase the CO
2
concentration in the operating
room.
Ensure sufficient supply of fresh air and use the smoke extraction function only if it
is necessary.
Fig.5–2Smoke evacuation indicator
(10) Increase smoke evacuation
level
(11) Set smoke evacuation level/
ON/OFF/PAUSE
(12) Reduce smoke evacuation level
(10)
(11)
(12)
An additional function of the device is the extraction of smoke from the cavity. Use
tubes with an additional lumen for smoke evacuation (see Chapter Insufflation
Tube Sets [}28]) to extract the smoke.
en