Users Manual

Operating the Device - General
29 / 77
SMOKE EVACUATION HIGH-FLOW TUBE
SET
(0620050250/black)
Disposable tube set with:
Filter
Lumen for smoke evacuation with
filter
To be used with the following operating
modes:
Standard, - Pediatric, - HighFlow/Bariat-
ric, - Advanced Flow, - TAMIS
HEATED HUMIDIFIED TUBE SET
(0620050300/red)
Disposable tube set with:
Filter
Gas heating
Gas humidification
To be used with the following operating
modes:
Standard, - Pediatric, - HighFlow/Bariat-
ric, - Advanced Flow, -Vessel Harvest, -
TAMIS
HEATED HUMIDIFIED SMOKE EVACU-
ATION TUBE SET
(0620050350/orange)
Disposable tube set with:
Filter
Gas heating
Gas humidification
Lumen for smoke evacuation with
filter
To be used with the following operating
modes:
Standard, - Pediatric, - HighFlow/Bariat-
ric, - Advanced Flow, -TAMIS
All tube sets are intended for single use.
The tube sets have been sterilized with ethylene oxide according to procedures val-
idated by ISO11135-1 and ISO10993-7.
RFID technology/Invalidating a tube set
Transponder technology (RFID) automatically detects the tube type, the validity, as
well as the reliability of a tube set.
After insertion of the tube set and starting insufflation, the tube set is marked as
used after 10minutes and thereafter can only be used with the same insufflator
and will not be valid for use with another insufflator. The tube set then can be used
for maximal 12hours before it is invalidated. However, for each patient, a new and
sterile tube set must be used. Insufflation can be stopped and restarted (start-stop-
button) during the use, this will not invalidate the tube set. The device may be
switched off and on for a limited time without invalidating the tube set. But if the
device is switched off for more than 2hours the tube set will be invalid and cannot
be used after restart.
If an invalidated tube set is connected to the device, the message
InvalidTubeSet
appears. If this is the case, insufflation is not possible unless a valid tube set is in-
serted.
5.3 Connecting an Insufflation Tube Set
WARNING!
Visual inspection of the tube set
Before the operation, perform a visual inspection of the tube set and its packaging.
Damaged tube sets or tube sets from damaged packagings may not be used.