Users Manual
Initial Device Setup
26 / 77
Icon Description
(red)
< 10 bar /145 psi - 4 bar /58 psi. Acoustic signals can be heard and a message is dis-
played. Replace the gas bottle.
(red)
< 4 bar /58 psi - 0 bar / 0 psi. Acoustic signals can be heard and a message is displayed.
Immediately replace the gas bottle.
Central gas supply
The following central gas supply icons are displayed:
Icon Description
(green)
Central gas supply pressure OK
(red)
Central gas supply pressure too low
4.6 Switching Device On
Fig.4–4Setting up the device
(1)
(2)
(3)
1. Connect the device with the mains power.
2. Connect the gas supply to the gas connection port and open the gas supply.
3. Press the ON/OFF switch. The power LED lights up green when the device is
turned on.
4. The device performs an automatic self-test after being switched on. If the self-
test is not successful, the corresponding error message will be displayed (see
Chapter Error and Warning Messages [
}65]).
5. The language selection is displayed when starting the device for the first time.
Select the desired language and confirm with the
Continue
button on the bot-
tom left.
6. The display depicts an overview of the insufflation operating modes. Unavail-
able operating modes are depicted in gray and cannot be selected. Press the re-
spective button to select the desired operating mode (e.g.
Standard
).
Unavailable insufflation operation modes can be unlocked. For this, contact your
customer service.
4.7 Switching Device Off
In order to safely terminate the operation of the system, press the ON/OFF switch.
Disconnect device from power supply
Disconnect the device from the mains power supply (pull cable out off the groun-
ded safety wall socket) if the device is not being used for several days or longer.