Users Manual
Initial Device Setup
21 / 77
4 Initial Device Setup
The operation of the device
PNEUMOCLEAR™
is reserved for medical staff with the
relevant professional qualifications trained to use the device.
NOTE!
Locations
The device may be used only in the hospital or operating rooms.
Delivery inspection
Always check all parts and optional accessories of the device immediately after re-
ceiving the shipment. The manufacturer only considers replacement claims that
have been immediately submitted or reported to a sales representative or an au-
thorized service company.
Returning the device
If it becomes necessary to return the device, always use the original packaging. The
manufacturer does not take responsibility for damage that has occurred during
transportation if the damage was caused by inadequate transport packaging.
Please keep the original packaging.
Please make sure that all required information has been supplied:
• Name of owner
• Address of owner
• Device type and model
• Serial number of the equipment (see identification plate)
• Detailed description of defect
4.1 Device Setup and Connection
Setting up
Place the device on a flat surface free of vibration located in a dry environment. The
ambient conditions must meet the requirements mentioned in ChapterTechnical
Data [}67].
CAUTION!
ME systems
The medical electrical (ME) device is suitable for integration in ME equipment sys-
tems (see Chapter Glossary). Operation of the ME device in vicinity of non-ME
devices may result in voiding the intended use of the ME device.
CAUTION!
Device setup
Device should be positioned outside of the sterile area in such a way that
• it can be easily disconnected,
• it is easy to use and switch off and on,
• it allows an easy monitoring of the display values, device functions, and access
to the control elements.
CAUTION!
Ventilation of the device
• Avoid device overheating.
• Ensure free air circulation especially to the bottom and rear of the device (rear
panel distance of at least 10cm).
en
DRAFT