Data Sheet

44
45
The blood sample
did not fill the
confirmation
window of the test
strip during
measurement
because of
abnormally high
viscosity or
insufficient volume.
Repeat test with
new test strip and
adequate blood
sample.
Message What It Means What To Do
If the error messages persist, contact your authorized i-SENS
sales representative.
Understanding Error and Other Messages
Message What It Means What To Do
Test strip has
been used.
Use a new strip.
Blood sample
applied before
the
symbol
appeared. Repeat
test with a new
strip.
Wait until the
symbol
appears before
applying sample.
Temperature
recommended
system range
during test.
Take system
where temperature
is within operating
range (10-40˚C/
50-104˚F ) and repeat
test after 30 minutes.