Data Sheet

40
41
Test Range: 20-600 mg/dL (1.1-33.3 mmol/L)•
Sample Size: Minimum 0.5 •
Test Time: 5 seconds•
Test Sample: Fresh capillary whole blood•
Calibration: Plasma-equivalent•
Assay Method: Electrochemical•
Meter Test: Check Strip•
Power: One (1) 3.0 V lithium battery•
(non-rechargeable, replaceable, type CR2032)•
Battery Life: 1,000 tests•
Memory: 250 blood glucose values•
Dimension: 95 x 41.5 x 17.5 (mm)•
Weight: 42.4 g (with battery)•
Operating Range•
- Temperature: 10-40˚C/50-104˚F
- Relative Humidity: 10-90%
- Hematocrit: 20-60%
Specifications
[1] Turn the meter over and at the top you
see the recess for openning the battery
compartment cover. Slide the cover out of
the meter.
[2] Remove the battery. Insert new battery
negative (-) side facing up.
[3] Slide cover down until it clicks.
Replacing the Batteries
When the icon appears on the display for the first time, the
battery should be replaced as soon as possible.
Use only non-rechargeable, replaceable, type CR2032 lithium
battery with 3 V nominal voltage, 220 mAh nominal capacity,
and a 0.2 mA nominal standard discharge.
There is a risk of explosion if the battery is replaced with an
incorrect type. Please dispose of used batteries according to
your local regulations.