Data Sheet
30
31
If confirmation window has not filled in time because of
abnormal viscosity or insufficient volume, Er4 message will
appear.
If the confirmation window is not completely filled when results
appear, it is recommended that you perform a second test for
more accurate results.
• Helpful Tip
If the test strip dose not receive a blood sample, please contact your
authorized i-SENS sales representative.
[3] Meter will count down from five-to-one
(5-1) on the display. Test results, time,
and date will appear and automatically
store in meter memory. Remove used
test strip from the port. Meter will turn off
after three (3) seconds.
Completely
filled
Correct Poorly
filled
Incorrect
[2] Use lancing device to get blood sample. Sample must be at
least 0.5 ㎕ (actual size: ) to fill the test strip reaction cell.
When the symbol appears on display, apply blood sample to
narrow top edge of test strip until meter beeps.
When testing from forearm, hold the meter as shown.
[1] Insert test strip with contact bars
facing up into a port. Push strip
gently until meter beeps. A code
number will appear for three (3)
seconds.
The symbol will appear.
Display code number should match
the one on the test strip bottle. If
not, see Coding Meter to Match
Test Strips on pages 19-20.
Testing Your Blood Glucose Levels