Repeat the control test if the test result falls outside the range
printed on label of the test strip vial. Out of range results may occur
due to the following factors:
When the control solution vial was not shaken well,•
When the control solution is past its expiration date or is •
contaminated,
When the meter, the strip or the control solution were •
exposed to high or low temperatures,
When the first few drops of the control solution were not •
discarded or the tip of the vial was not wiped clean,
When the test strip is past its expiration date,•
When the meter is not functioning properly.•
Comparing the Control Solution Test Results
The CareSens Control Solution can be purchased separately.
Please contact your authorized i-SENS sales representative.
[3] Test results appear after the meter
counts down from five-to-one (5-to-1).
To mark test result, press and hold the
C Button for three (3) seconds. Check
will appear on the display. Marked test
results will not be shown in your
averages.
[4] Compare results with range printed on
test strip bottle. If results fall within range,
you are ready to test your blood.
[5] Remove used test strip. The Meter will
turn off.