Data Sheet

22
23
[1] Insert test strip contact bars facing
up into the meter’s test strip port. Gently
push test strip into the port until meter
beeps. A code number will appear for
three (3) seconds.
The symbol will show on the display.
Confirm matching code numbers on
display and test strip bottle. If code
numbers do not match, see Coding Meter
to Match Test Strips on pages 19-20.
[2] Shake the bottle well.
Discard first drop of the
control solution.
Wipe off the bottle tip.
After the symbol
appears on the display,
apply the solution to
narrow top edge of the
test strip until the meter
beeps.
Perform the following steps to check the combined performance
of your CareSens II Meter and Test Strips or to assess proper testing
procedure performance.
The glucose level printed on the bottle is for the CareSens
Control Solution only.
It does not refer to your blood glucose level.
Testing Meter and Test Strip Performance
Discard the CareSens Control Solution after its expiration date
on the bottle or discard date, whichever comes first. Add three
(3) months to the date you open the bottle and write it on the
bottle label. This is the discard date. Do not use the solution
after the discard or expiration date, whichever comes first.
For accurate results:
Store control solution in tightly capped bottle at •
temperature 8-30°C /46-86°F.
Do not refrigerate or freeze.•
Keep the control solution, meter, and test strips at room •
temperature (20-25°C/68-77°F) for at least 30 minutes
before testing.
Do not use the control solution after the expiration or
discard date, whichever comes first. Only use CareSens
Control Solution to test CareSens Test Strips and the
CareSens II Meter.
CareSens Gloucose Control Solution
Storage and Handling
If all the results you get on testing three times continuously
are within the range printed on the test strip, the meter and
test strips are working properly and you may use them for
your blood glucose test.