Data Sheet

Electrocardiography (ECG)
Sensor Data Sheet
!
PAGE 2 OF 2
!
!
TRANSFER FUNCTION
[-1.5π‘šπ‘‰, 1.5π‘šπ‘‰]
𝐸𝐢𝐺 𝑉 =
𝐴𝐷𝐢
2
!
βˆ’
1
2
. 𝑉𝐢𝐢
𝐺
!"#
𝐸𝐢𝐺 π‘šπ‘‰ = 𝐸𝐢𝐺 𝑉 . 1000
𝑉𝐢𝐢 = 3.3𝑉 (operating voltage)
𝐺
!"#
= 1100 (sensor gain)
𝐸𝐢𝐺 𝑉 – ECG value in Volt (𝑉)
𝐸𝐢𝐺 π‘šπ‘‰ – ECG value in millivolt (π‘šπ‘‰)
𝐴𝐷𝐢 – Value sampled from the channel
𝑛 – Number of bits of the channel
1
!!!!!!!!!!!!!!!!!!!!!!!!!!!!!!!!!!!!!!!!!!!!!!!!!!!!!!!!
1
!The number of bits for each channel depends on the resolution of the Analog-to-Digital
Converter (ADC); in BITalino the first four channels are sampled using 10-bit resolution
(𝑛 = 10), while the last two are sampled using 6-bit (𝑛 = 6).