4.2 Synchronization & Additional Digital Ports
4.2.1 Digital Synchronization Cable
Figure 79: Sync cable & kit.
The multi sync splitter accessory meets your multi-device acquisition needs. By using
digital synchronization cable(s), this accessory allows you to connect biosignalsplux
devices with digital ports together to collect up to 24 channels at the same time (by
connecting up to 3 HUBs together).
NOTE
If you are using more than 2 devices, follow the instructions provided in 4.2.2 Multi Sync
Splitter).
How to connect your accessory to your biosignalsplux
The digital synchronization cable needs to be connected to the digital ports of the devices
that are being used for the multi-device acquisition (see 2.3.4 Digital Port).
How to configure synchronized acquisitions in OpenSignals (r)evolution
Open the OpenSignals (r)evolution device manager and select one of the devices you want
to use for the multi-device acquisition. In the device’s panel, click on the circle next to
SYNC to turn on (blue circle) or off (empty circle) the synchronization function to
synchronize the acquisition of your devices (see Figure 80). This function must be active
to ensure synchronized multi-device acquisitions without signal drifts. Repeat this step
for the second device that is being used for the multi-device acquisition.