should be showing Pc”) and the signal will now be adjusted according to the last
measured value.
Error Codes
The sensor will provide error codes if measurements could not be made correctly.
Depending on the type of error, the sensor will display one of the different error codes
below:
Error Code
Issue
Solution
Er1
The test strip you are trying to use
has already been used.
Dispose the used test strip and use a
new test strip. Dispose the used lancet
and use a new one for the new
measurement.
Er2
The blood sample has been applied
before the sensor has been ready
for measurements.
Dispose the used test strip and Use a
new test strip and wait for the lancet
symbol to appear at the bottom of the
display before applying your blood
sample. Dispose the used lancet and use
a new one for the new measurement.
Er3
Temperature recommended system
range during test.
Take the system where the temperature
is within the operating range (10-40°)
and repeat test after 30 minutes.
Er4
The blood sample has an
abnormally high viscosity or
insufficient volume.
Repeat the test with a new test strip and
adequate blood sample. Dispose the
used lancet and use a new one for the
new measurement.
If required, increase the depth of the
lancet and massage the fingertip before
taking the blood sample, to increase the
size of the blood sample.
Table 5: Glucose meter reader error codes.