Data Sheet

biosignalsplux
User Manual
!
!
!
76!of!159!
!
3.2.10 SpO2 Reader (OSL)
Figure 64: SpO2 reader.
This comfortable and user-friendly finger clip sensor was designed to measure reliable
oxygen saturation levels (%) and heart rate (bpm) information. The built-in colour LCD
display displays all measured values and signals and allows this sensor to be used both
as standalone device or as biosignalsplux sensor within our OpenSignals (r)evolution
software.
NOTE
This sensor is an adaptation of the 3rd party CMS-50D Plus which provides compatibility
with biosignalsplux devices. It can be used both as standalone device or as
biosignalsplux sensor.
Please do also read the datasheet of the standalone device before using this sensor:
http://biosignalsplux.com/datasheets/telemedicine/CMS-50D_Plus.pdf
How to connect your sensor to your biosignalsplux
Connect the sensor to the adapter box using the micro-USB cable which comes pre-
connected to the adapter box. When done, connect the outgoing cables to any of the
available analog inputs of your biosignalsplux (see 2.3.2 Analog Inputs). Note that this
sensor has individual channels for the SpO2 and heart rate signals.
How to configure your sensor in OpenSignals (r)evolution
Open the OpenSignals (r)evolution device manager and select the device you have your
OSL sensor connected to. Select the channel of your SpO2 signal and set the sensor type