Figure 59: biosignalsplux device panel with firmware 3.4. 
How to connect your sensor to your biosignalsplux 
The fNIRS sensor needs to be connected to the ground port of your device. Connecting 
this sensor to an analog input of your device will not work. 
How to configure your sensor in OpenSignals (r)evolution 
Open the OpenSignals (r)evolution device manager and select the device you have your 
fNIRS sensor connected to. 
In the device’s  panel,  set  up  the  channels  9  and  10  for  signal  acquisition  and  set  the 
sensor type to fSpO2 or, if not available, to SpO2.HEAD (click on the arrow in the channel’s 
field that appears when you hover over the channel’s field). For channel 9, set the port 
(P) to 9 and the channel (C) to 1. For channel 10, set the port (P) to 9 and the channel to 
(C) 2. 
Channel 9 visualizes the reflected light of the red LED, while channel 10 visualizes the 
reflected light of the IR. The intensities of the LEDs can be configured individually by 
changing  the  sliders  at  the  left  side  next  to  the  channel  configuration  panels.  If 
configured correctly, the device panel should be similar to Figure 60. 
Figure 60: Configuration of the fNIRS sensor with biosignalsplux. 