9. Safety, Care, Accuracy Test and Disposal
Safety and protection
«
ERR 2
» Error signal During the measurement, error signals were
detected by the cuff, caused for instance by
movement or muscle tension. Repeat the
measurement, keeping your arm still.
«
ERR 3
» No pressure
in the cuff
An adequate pressure cannot be generated
in the cuff. A leak may have occurred. Check
that the cuff is correctly connected and is
not too loose. Replace the batteries if neces-
sary. Repeat the measurement.
«
ERR 5
» Abnormal
result
The measuring signals are inaccurate and no
result can therefore be displayed. Read
through the checklist for performing reliable
measurements and then repeat the
measurement.*
«
HI
» Pulse or cuff
pressure too
high
The pressure in the cuff is too high (over
300 mmHg) OR the pulse is too high (over
200 beats per minute). Relax for 5 minutes
and repeat the measurement.*
«
LO
» Pulse too low The pulse is too low (less than 40 beats per
minute). Repeat the measurement.*
Error Description Potential cause and remedy
)
If you think the results are unusual, please read through the
information in «Section 1.» carefully.
•
This instrument may be used only for the purpose described in this
booklet. The manufacturer cannot be held liable for damage
caused by incorrect application.
•
This instrument comprises sensitive components and must be
treated with caution. Observe the storage and operating condi-
tions described in the «Technical Specifications» section!
•
Protect it from:
− water and moisture
− extreme temperatures
− impact and dropping
− contamination and dust
− direct sunlight
− heat and cold
•
The cuffs are sensitive and must be handled with care.