User manual

Menus48
Copyright © 2007-2013 Pico Technology Ltd. All rights reserved.psw.en r32
Filter control
PicoScope can low-pass filter the statistics to produce more
stable and more accurate numbers. Filtering is not available on
all measurement types.
Enable Filter - check to enable low-pass filtering, if available.
An "F" will appear after the measurement name in the
measurements table.
Automatic - check to set the low-pass filter characteristics
automatically
Cutoff Frequency
The filter cut-off frequency normalised to the measurement
rate. Range: 0 to 0.5.
Filter Size
The number of samples used to construct the filter
Harmonic Control
These options apply to distortion measurements in spectrum
views. You can specify which harmonics PicoScope uses for
these measurements.
Highest Harmonic
The highest harmonic to include when calculating distortion
power
Search Range
The number of frequency bins to search, centred on the
expected frequency, when looking for a harmonic peak
Harmonic Noise
Floor
The level in dB above which signal peaks will be counted as
harmonics