PicoScope 6 User's Guide 129
Copyright © 2007-2013 Pico Technology Ltd. All rights reserved. psw.en r32
7.7.1.1
Basic controls
Signal generator dialog for the
PicoScope5204
Signal On. Tick this box to enable the signal generator.
Signal Type. Select the type of signal to be generated. The list of signal types
depends on the capabilities of the scope device.
Import. Opens a file selection dialog that allows you to import an arbitrary waveform
file. The file will be loaded into the arbitrary waveform generator and the generator
switched on. This button is available only if your scope has an arbitrary waveform
generator.
Arbitrary. Opens the Arbitrary Waveform window. This button is available only if
your scope has an arbitrary waveform generator.
Start Frequency. Type in this box or use the spin buttons to select the frequency. If
the scope device has a frequency sweep generator, then this box sets the start
frequency of the sweep.
Amplitude. The amplitude of the waveform measured from peak to peak. For
example, if Amplitude is 1 V and Offset is 0 V, the output will have a negative peak
of –0.5 V and a positive peak of +0.5 V.
Offset. The mean value of the signal. For example, when Offset is 0 V, a sine or
square wave will have equal positive and negative peak voltages.
7.7.1.2
Sweep controls
The signal generator normally generates a fixed frequency set by the Start
Frequency Control. In sweep mode, it generates a frequency that varies between
two specified limits.
Active. Tick this box to enable sweep mode.
Sweep Type. Specifies the direction in which the frequency is swept.