User manual

PicoScope 6 User's Guide 123
Copyright © 2007-2013 Pico Technology Ltd. All rights reserved. psw.en r32
Average: the spectrum view shows a rolling average of spectra
calculated from all the waveforms in the waveform buffer. This
has the effect of reducing the noise visible in the spectrum view.
To clear the averaged data, click Stop and then Start, or
change from Average mode to Magnitude mode.
Peak Hold: the spectrum view shows a rolling maximum of the
spectra calculated from all the waveforms in the buffer. In this
mode, the amplitude of any frequency band in the spectrum
view will either stay the same or increase, but never decrease,
over time. To clear the peak hold data, click Stop and then
Start, or change from Peak Hold mode to Magnitude mode.
Note: when you switch to Average or Peak Hold mode, there
may be a noticeable delay while PicoScope processes the entire
contents of the waveform buffer, which may contain many
waveforms, to build up the initial display. If this occurs, a
progress bar appears at the bottom of the window to show that
PicoScope is busy:
Scale
Specifies the labelling and scaling of the vertical (signal) axis.
This can be one of the following:
Linear:
The vertical axis is scaled in volts.
Logarithmic:
The vertical axis is scaled in decibels, referred to the level
selected below in the Logarithmic unit control.
dBV: Reference level is 1 volt.
dBu: Reference level is 1 milliwatt with a load resistance of
600 ohms. This corresponds to a voltage of about 775 mV.
dBm: Reference level is one milliwatt into the specified load
impedance. You can enter the load impedance in the box
below the Logarithmic unit control.
Arbitrary dB: Reference level is an arbitrary voltage, which
you can specify in the box below the Logarithmic unit
control.