Logic control
Specifies the Boolean operation used to combine the input
trigger conditions. Only inputs with the 'Used' box checked
(see above) are included in the trigger logic.
AND: all of the input trigger conditions must be met
NAND: none of the input trigger conditions must be met
OR: one or more of the input trigger conditions must be
met
NOR: none of the input trigger conditions must be met
XOR: exactly one of the input trigger conditions must be
met
XNOR: all but one of the input trigger conditions must be
met