Datasheet

PicoScope
®
5000D Series oscilloscopes
PicoScope 5000D Series
Technical Specications
PicoScope 5242D
and 5242D MSO
2-channel, 60 MHz
PicoScope 5243D
and 5243D MSO
2-channel, 100 MHz
PicoScope 5244D
and 5244D MSO
2-channel, 200 MHz
PicoScope 5442D
and 5442D MSO
4-channel, 60 MHz
PicoScope 5443D
and 5443D MSO
4-channel, 100 MHz
PicoScope 5444D
and 5444D MSO
4-channel, 200 MHz
Vertical (digital channels) – D MSO models only
Input channels 16 channels (2 ports of 8 channels each)
Input connector 2.54 mm pitch, 10 x 2 way connector
Maximum input frequency 100 MHz (200 Mbit/s)
Minimum detectable pulse width 5 ns
Input impedance 200 kΩ ±2% || 8 pF ±2 pF
Input dynamic range ±20 V
Threshold range ±5 V
Threshold grouping Two independent threshold controls. Port 0: D0 to D7, Port 1: D8 to D15
Threshold selection TTL, CMOS, ECL, PECL, user-dened
Threshold accuracy < ±350 mV including hysteresis
Threshold hysteresis < ±250 mV
Minimum input voltage swing 500 mV peak to peak
Channel-to-channel skew 2 ns, typical
Minimum input slew rate 10 V/µs
Overvoltage protection ±100 V (DC + AC peak)
Horizontal
Max. sampling rate
Any 1 channel*
Any 2 channels*
Any 3 or 4 channels*
More than 4 channels*
8-bit mode
1 GS/s
500 MS/s
250 MS/s
125 MS/s
12-bit mode
500 MS/s
250 MS/s
125 MS/s
62.5 MS/s
14-bit mode
125 MS/s
125 MS/s
125 MS/s
62.5 MS/s
15-bit mode**
125 MS/s
125 MS/s
16-bit mode**
62.5 MS/s
*”Channel” means any analog channel or 8-bit digital port
**Any number of 8-bit digital ports can be used in 15-bit and 16-bit modes without affecting the maximum sampling rate.
Maximum equivalent sampling rate
(repetitive signals; 8-bit mode only)
2.5 GS/s 5 GS/s 10 GS/s 2.5 GS/s 5 GS/s 10 GS/s
Maximum sampling rate
(continuous USB streaming into PC
memory)
USB3, using PicoScope 6: 15 to 20 MS/s (split between enabled channels, PC dependent, available sample rates vary by resolution)
USB3, using API: 125 MS/s (8-bit) or 62.5 MS/s (12 to 16 bit) (split between enabled channels, PC dependent)
USB2, using PicoScope 6: 8 to 10 MS/s (split between enabled channels, PC dependent, available sample rates vary by resolution)
USB2, using API: ~30 MS/s (8-bit) or ~15 MS/s (12 to 16 bit) (split between enabled channels, PC dependent)
USB1.1 (not recommended) :~1 MS/s (8-bit) or ~0.5 MS/s (12 to 16 bit) using PicoScope 6 or supplied API (split between enabled channels,
PC dependent)
Timebase ranges (real time) 1 ns/div to 5000 s/div in 39 ranges
DRAFT