Datasheet
PicoScope
3203D and
3203D MSO
PicoScope
3403D and
3403D MSO
PicoScope
3204D and
3204D MSO
PicoScope
3404D and
3404D MSO
PicoScope
3205D and
3205D MSO
PicoScope
3405D and
3405D MSO
PicoScope
3206D and
3206D MSO
PicoScope
3406D and
3406D MSO
Triggering
Source
Analog channels (all models)
EXT trigger (D models only)
Digital channels (D MSO models only)
Trigger modes None, auto, repeat, single, rapid (segmented memory)
Maximum pre-trigger
capture
Up to 100% of capture size
Maximum post-trigger delay Up to 4 billion samples, selectable in 1 sample steps
Trigger rearm time < 0.7 µs at 1 GS/s sampling rate
Maximum trigger rate Up to 10 000 waveforms in a 6 ms burst at 1 GS/s sampling rate, typical
Triggering for analog channels
Advanced trigger types Edge, window, pulse width, window pulse width, dropout, window dropout, interval, logic, runt pulse
Trigger types (ETS mode) Rising edge, falling edge (available on channel A only)
Trigger sensitivity Digital triggering provides 1 LSB accuracy up to full bandwidth of scope
Trigger sensitivity
(ETS mode)
10 mV p-p at full bandwidth, typical
External trigger input - D models only
Ext trigger connector type Front panel BNC
Trigger types Edge, pulse width, dropout, interval, logic
Input characteristics 1MΩ
∥ 14 pF
Bandwidth 50 MHz 70 MHz 100 MHz 200 MHz
Threshold range ±5 V
Coupling DC
Overvoltage protection ±100 V (DC + AC peak)
Triggering for digital channels - D MSO models only
Trigger types Pattern, edge, combined pattern and edge, pulse width, dropout, interval, logic