Manual
4.0 SPECIFICATIONS
4.1 CONTROLLER
Operating Range: -20
o
to +60
o
C
Control Accuracy:
+
0.1
o
C
Digital Readout Resolution: 0.1
o
C
Accuracy: 0.1
o
C,
+
1 digit
Ambient Operating Range: 15
o
- 45
o
C
Input Power Requirements: 100 - 120VAC, 60Hz, 100W
200 - 240VAC, 50Hz, 100W
Safety Features: Safety shutdown with warning lamp in case of fault condition
such as sensor breakage, lack of cooling water or electronics
failure.
Size: 8” high x 7” wide x 15” deep
Weight 8 lbs.
Other Features: Spring loaded switch indicates set point
Self-sealing water connectors
Auxiliary AC switched output
4.2 STAGE
Temperature Range: -20
o
to +60
o
C
Controlled Surface Area: 1 1/4” x 1 1/2”
Weight: 16 ozs., including leads
Material: Copper, nickel plated
Mounting: Plastic. Fits slide holder
4.3 PUMP AND TANK UNIT
Size: 6 1/2” x 15” x 14”
Power Supply: 110V (or 220VAC if ordered)
Power Consumption: 150 Watts
Water Connections: Self Sealing
4.4 THERMOMETER (built into controller)
Temperature Range: -100
o
to +200
o
C
Resolution: 0.1
o
C
Sensor: Type T Thermocouple, Physitemp type #MT-26/4 or similar
7