OPERATING MANUAL CT-3 TEMPERATURE STIMULATOR CONTENTS Section 1.0 GENERAL DESCRIPTION 2.0 INITIAL INSTALLATION 3.0 OPERATING INSTRUCTIONS 4.0 BATTERY REPLACEMENT 5.0 SPECIFICATIONS 6.0 CALIBRATION 7.
CT-3 OPERATING MANUAL 1.0 GENERAL DESCRIPTION 1.1 The CT-3 is a simple device that simulates thermocouple signals so that the accuracy of a thermocouple thermometer can be checked instantly, without waterbaths. 1.2 The regular CT-3 is designed for use with type T thermocouple thermometers and has fixed electrical signals corresponding to five temperatures: 0.0 , 15.0 , 37.0 , 45.0 and 50.0 C. It is possible to order CT-3 with other settings.
thermometer under test, continue the calibration check. If they do not, the thermocouple thermometer may require recalibration. Note: ensure that the mercury thermometer is viewed perpendicular to the tip of the mercury column to avoid parallax effects. 3.4 Press the grey ON ADJUST button on the CT-3. While holding down this button, turn the ADJUST knob until the thermocouple thermometer displays 25.0 C (77.0 F) 3.5 Continue to hold down the grey button. Press in turn the 0.0 , 15.0 , 37.0 , 45.
5.0 SPECIFICATIONS Thermocouple Type: Type T, copper-constantan Temperature Simulated: 0.0 , 15.0 , 25.0 , 37.0 , 45.0 , 50.0 C Accuracy: Better than 0.1 C Ambient Operating Temperature: 15 to 35 C (Mercury-in-glass thermometer) Power Supply: 2 “N” size alkaline cells, 1.5v each Expected battery life: 200 hours continuous operation. 2 years intermittent Low Battery warning: LED Size: 11 3/16” L x 3 9/16”W x 2 3/8” H Weight: 24 0z. 6.
7.0 WARRANTY AND SERVICE 7.1 WARRANTY Phystiemp warrants this instrument to be free from defects in material or workmanship for 12 months from date of shipment. Repair or replacement will be made at no charge at the discretion of Phystimep if the defect is not the result if misuse or abuse. Physitemp accepts no consequential liability for delay in delivery, alleged faulty performance of the product or any other cause. 7.