User guide
18 TrueCPR Device Instructions for Use
Product Specifications
Product Specifications
Safety Classifications
SPECIFICATION DESCRIPTION
Power Requirements Batteries: 2 Duracell nonrechargeable DL123 cells
Operating Time 180 minutes at room temperature with new batteries
Operating Temperature Device can be used at 0° to 40°C (32° to 104°F).
Applied part temperature may reach 43°C (109°F) during
extended use at ambient temperature of 40°C (104°F).
Shelf Life Provides 30 minutes of CPR feedback after 24 months without
use at 25°C (77°F), starting with new batteries installed.
Storage Temperature -30° to 70°C (-22° to 158°F)
Relative Humidity 5 to 90%, non-condensing
Altitude -382 to 4,572 m (-1,253 to 15,000 ft)
Object and Water
Resistance
IP55 per IEC 60529 and EN 1789
Weight Less than 0.75 kg (1.65 lb) with batteries installed
Ventilation Prompts 2 ventilation prompts every 30 compressions in No Airway mode.
No ventilation prompts in Airway mode.
Compression Depth Target depth range of 5 to 6 cm (2 to 2.5 in)
Metronome Rate 104.4
± 1 compressions per minute, consistent with AHA and
ERC guidelines.
Model 80596-000001
CATEGORY CLASSIFICATION
Protection Against Electric Shock Internally powered ME equipment.
Type BF defibrillation proof applied parts.*
Protection Against Solid Object and
Liquid Ingress
IP55
Mode of Operation Continuous operation
* Type BF defibrillation proof applied parts are the underside of the chest pad and the top
of the back pad. See page 17 for symbol illustration.
3308456-900.book Page 18 Wednesday, October 24, 2012 9:51 AM